Mother and son jogging Mother and son jogging Mother and son jogging Mother and son jogging
Abbott Implements Corrective Action for HeartMate 3 Heart Pump
- Abbott previously communicated safety alert related to potential outflow graft twisting to physicians on April 5, 2018

ABBOTT PARK, Ill., May 4, 2018 -- Abbott is communicating to physicians about a field corrective action related to the HeartMate 3 Left Ventricular Assist Device (LVAD). Abbott notified physicians and global regulatory bodies on April 5, 2018, that outflow graft twisting could occur post-implant in the HeartMate 3 LVAD, resulting in a persistent low flow alarm that may signal a potential safety risk to patients, such as low blood flow or clotting.

No devices are being recalled from patients or hospitals as a result of the corrective action, and the recommendations made by Abbott remain in place for physicians managing patients implanted with HeartMate 3.

Abbott is currently aware of 32 reports of outflow graft twisting in 4,467 HeartMate 3 devices worldwide (incidence rate of 0.72%), including reports of low blood flow, clotting (known as thrombus), and three reported deaths that could be associated with outflow graft twisting. As a result, Abbott has contacted physicians managing patients implanted with a HeartMate 3 LVAD in the event their patients report persistent low flow alarms that could indicate outflow graft twisting. Twisting of the outflow graft can occur at any point after implant.

In patients living with an LVAD, low flow alarms can often indicate unrelated conditions such as hypertension. For patients who experience a persistent low flow alarm after such causes have been ruled out, a computed tomography (CT) scan can help detect any possible outflow graft twisting. In addition, physicians managing patients that exhibit a persistent low flow alarm should determine patient care recommendations based on each unique clinical case.

Abbott has also alerted implanting physicians that as stated in the device’s instructions for use (IFU), properly hand tightening the screw ring at the time of implant may reduce the chance of outflow graft twisting. Abbott has proactively contacted all physicians who treat patients impacted by this safety alert and will continue to assess current reports of outflow graft twisting to further provide physicians with patient care recommendations.

About Abbott 
At Abbott, we're committed to helping you live your best possible life through the power of health. For more than 125 years, we've brought new products and technologies to the world – in nutrition, diagnostics, medical devices and branded generic pharmaceuticals – that create more possibilities for more people at all stages of life. Today, 99,000 of us are working to help people live not just longer, but better, in the more than 150 countries we serve.

Connect with us at www.abbott.com, on Facebook at www.facebook.com/Abbott and on Twitter @AbbottNews and @AbbottGlobal. 

For further information: Abbott Media: Justin Paquette, 651-756-6293; Abbott Financial: Mike Comilla, 224-668-1872
true
accessibility

You are about to exit for another Abbott country or region specific website

Please be aware that the website you have requested is intended for the residents of a particular country or countries, as noted on that site. As a result, the site may contain information on pharmaceuticals, medical devices and other products or uses of those products that are not approved in other countries or regions.


The website you have requested also may not be optimized for your specific screen size.

Do you wish to continue and exit this website?

accessibility

You are about to exit the Abbott family of websites for a 3rd party website

Links which take you out of Abbott worldwide websites are not under the control of Abbott, and Abbott is not responsible for the contents of any such site or any further links from such site. Abbott is providing these links to you only as a convenience, and the inclusion of any link does not imply endorsement of the linked site by Abbott.


The website that you have requested also may not be optimized for your screen size.

Do you wish to continue and exit this website?